Product Info :: Articulation Durability

Goal

Ensure that wear debris is minimized.

Conclusion

TFAS® articulating surfaces produced very small amounts of wear; comparable to other successful orthopaedic implants. The debris that was produced caused no adverse reactions in surrounding tissues.

Articulation Wear Performance

  • Tested per ASTM F-1717 (modified)
  • Fatigue testing to 10 million cycles
  • Supports 2-3x maximum static in vivo loads
  • Supports = 2x maximum fatigue in vivo loads
  • Integrity of the construct & interconnection mechanisms at 10-million cycles
  • No dislocation or failure under maximum static loading